SentroPAT DA
Diode array NIR analyzer for pharmaceutical powder blending | |
|
Main features
- Lightweight, compact, easy to mount
- GmP friendly with smooth surfaces
- Accurate and reliable Diode Array NIR technology
- Internal wavelength and intensity reference
- Lowest cost of ownership
- WiFi connected and battery powered
- Embedded PC for instrument control and data buffering
- Full GmP support for instrument and
- Optional Interfaces to OPC, SiPAT, SynTQ or PharmaMV
|
| Bin Blending Sentronic PAT solutions are widely used in pharmaceutical powder blending to increase process understanding and support process and product release |
- Fits easily even to small scale laboratory blenders
- High spectral resolution and low noise for high analytical quality
- Internal intensity and wavelength references provide stability and ease of use
- Cost efficient solution
- Process interface compatibel to SentroPAT BU II when using the 4" flange
|
Technical Data | SentroPAT DA |
---|
Detector | 512 pixel linear InGaAs array |
Wavelength range | 1100 - 1900nm |
Spectral resolution (FWHM) | approx. 10nm |
Wavelength accuracy | < 1nm |
Wavelength reproducibility | < 0.2nm |
Photometric linearity | Fulfilling USP <1856> and EP 2.2.40 requirements |
Noise - High flux | According to USP<1856> and E.P. 2.2.40 |
Noise - Low flux | According to USP<1856> and E.P. 2.2.40 |
Light source | 2 x 5W used redundantly for maintenance free operation for > 10 years |
Data acquisition time | < 1s (configurable) |
Sampling interface | Non contact, through sapphire window |
Measurement spot size | 12 mm |
Power supply | replaceable Li-Ion battery, > 6 hours operation External power supply for lab operation and recharging |
Enclosure | IP65 / NEMA4, cGMP conform, Aluminum anodized |
Dimensions | 170 x 105 x 60mm (W x D x H) Height: add 36mm for flange |
Weight | 2.2Kg |
Measurement trigger | software controlled by 3 axis accelerometer |
Communications | WiFi (802.11 a/g/n) |
Software | SentroSuite GmP, optional interfaces to SiPAT and SynTQ |
Validation | Software integrated validation routines according USP<1856>, EP2.2.40 |
Service port | Service Connections for HDMI monitor and USB for computer peripherals |
Mounting | 4" or 2" sanitary flange |